Sterilizer tests
Place the tube with the capsule in the center of a tray in the lowest rack position
4
and close the chamber door.
On the homepage, tap Helix-B&D test.
5
To set the duration of the Plateau/Sterilization phase and other settings, tap
6
Tap
and enter your credentials if required: the chamber door locks.
7
Wait until the end of the test and tap OPEN: the chamber door unlocks.
8
Enter your credentials if required.
9
Open the chamber door, extract the tray using the tray holder and take the tube.
10
CAUTION! Risk of burns. The tube is very hot at the end of the cycle. Wear
appropriate PPE (e.g. gloves).
Unscrew the tube capsule and remove the indicator strip.
11
Check the change in colour. See "Interpret the test result" below
12
INTERPRET THE TEST RESULT
Indicator
WHAT TO DO NEXT
Compile the Helix test documentation form to trace the effectiveness of the sterilization cycle during the whole lifespan of your sterilizer. See
"Helix test documentation form" on page  119.
44
What happened
The indicator strip has turned dark. The air was completely removed from
the capsule.
Part of the chemical indicator strip has not turned dark. The air removal
from the capsule was not complete.
.
Test passed What to do next
Yes
No
Lyla  | Instructions for Use |  RIN-210 ENG Rev02 |  15/09/2022 |  © W&H Sterilization Srl
-
Repeat the test. If it fails repeatedly,
call technical service.