Ordinary maintenance

MAINTENANCE BY THE USER
1
1
Frequency
Cycles
Operation
Cleaning the door gasket and the chamber face
Monthly
50
side. See "Cleaning the door gasket and the
chamber face side" on page  63
Clean the chamber, steam diffuser plate, trays
and the rack. See "Cleaning the chamber and the
chamber accessories" on page  64
Cleaning the chamber filter. See "Cleaning the
chamber filter" on page  65
Cleaning the external surfaces of the sterilizer.
See "Cleaning the external surfaces of the
sterilizer" on page  67
6 month
800
Clean both water tanks. See "800-cycle or
biannual maintenance" on page  71.
2
2
Replace the bacteriological filter. See "400-cycle
Yearly
400
maintenance" on page  68.
Replace the dust  filter. See "400-cycle
maintenance" on page  68
2
2
Replace the door gasket. See "800-cycle
Yearly
800
maintenance" on page  77.
5 years
4000
General check and service. See "4000 cycle or
five-year maintenance" on page  79.
1
: whichever occurs first.
2
: even if the maximum cycle number is not reached, it is recommended to
replace the consumable parts every year, or if they appear worn or
Lyla  | Instructions for Use |  RIN-210 ENG Rev02 |  15/09/2022 |  © W&H Sterilization Srl
damaged, or if the filters are clogged or discolored.
EXPIRED MAINTENANCE
The sterilizer monitors the wear of consumables by counting the
number of cycles executed since the last replacement.
When the number of cycles is close to the maximum, a pre-alert
about the concerned consumable is displayed. Please check that
you have the requested spare part available, buy one if not. When
the maximum number of cycles has been met, a message to replace
the consumable will be displayed.
If you can not replace the consumable immediately, the sterilizer
will operate anyway but the message will appear again some cycles
later.
When you have replaced the consumable tap
executed cycle counter is reset.
REPLACE THE CONSUMABLE BEFORE THE MAINTENANCE DUE
DATE
If you replace the consumables before the request of replacement
appears, you should manually reset the counters through the
following procedure.
On the homepage, tap
1
Select the consumable you want to replace: a message appears
2
showing the current worked hours of the part.
to confirm: the
>
Maintenance
61