Relay Card Connections - Hach sc1000 User Manual

Hide thumbs Also See for sc1000:
Table of Contents

Advertisement

Installation
Figure 20
Expansion Card Ports
10
1.
Relay Card
2.
Relay wiring information
3.
Field-bus or mA Output or Input Card
4.
Field-bus or mA Output or Input card wiring information
5.
mA Output or Input Card

3.6.1 Relay Card Connections

网址:www.mamots.com
24
1
2
3
4
Relay Module YAB022
PROFIBUS-DP YAB020
1
5V
Relay 1
2
3
B
R
IN
SPST-NC
3
5
S2
B
OUT
250VAC
5A
Relay 2
4
R
125VAC
5A
4
A
5
IN
125VDC
0,15A
R
6
S2
A
OUT
Relay 3
30VDC
5A
6
7
7
SC 1000
Relay 4
8
BUS
YAB020
PROFIBUS-DP
0000000
YAB022
Relay
0001439
1 2 3 4 5 6 7 8 9
DANGER
Relays must either be wired as low or high voltage.
DANGER
Relay loads must be resistive. User must externally limit current to the relays to
5 Amps by use of a fuse or breaker.
The relay connector accepts 18–12 AWG wire (as determined by load application). Wire
gauge less than 18 AWG is not recommended.
If the instrument is equipped with the relay card option, the instrument will include 4
relays, each with one normally closed contact. The relays can switch maximum 250 VAC,
5A (normally closed). Each relay can be configured for limits, status, or timer.
Refer to the Plug-in Expansion Card Instruction Sheet for information on installing the
relay plug-in expansion card.
1. Remove power from the instrument. Remove the probe module cover.
2. Remove the screws on the plastic relay cover. Remove the plastic cover
(Item 10,
Figure
20).
5
6
7
8
Input Module YAB018
Output Module YAB019
Analog or Digital
R=180
1
+
0...20mA
Analog
Analog
1
1
2
+
+5 V
max 500
-
Output 1
Output 1
JP
0...20mA
R
L H
R
-
2
2
3
+
0...20mA
Analog
Analog
3
3
max 500
4
Output 2
Output 2
+
+5 V
-
JP
0...20mA
R
L H
R
-
4
4
5
+
0...20mA
Analog
5
5
max 500
6
Output 3
+
+5 V
-
JP
0...20mA
R
L H
R
-
6
6
7
+
0...20mA
Analog
7
7
+5 V
max 500
8
Output 4
+
-
JP
0...20mA
R
L H
8
R
-
8
9
9
9
YAB018
YAB019
Input
Output
0000540
0000345
1 2 3 4 5 6 7 8 9
1 2 3 4 5 6 7 8 9
6.
mA Output or Input wiring information
7.
mA Output or Input Card
8.
mA Output or Input wiring information
9.
Main High Voltage Barrier
10. Relay Voltage Barrier
客服:400 688 9809
9
电话:0431-85178766

Advertisement

Table of Contents
loading

Table of Contents