Installing Power Outlets - Hach sc1000 User Manual

Hide thumbs Also See for sc1000:
Table of Contents

Advertisement

Terminal Number
L
N
Figure 15
Inside the AC Probe Module
13
12
1.
Fan
2.
Main circuit board
3.
Connector for expansion slot #2
4.
Connector for expansion slot #3
5.
Connector for expansion slot #4
6.
Fuse (2), F1 and F2: M 3.5A, medium slow-low
7.
Fuse (2), F3 and F4: T 8A; 100–240 V

3.4.4 Installing Power Outlets

网址:www.mamots.com
Table 5 AC Power Wiring Information
Terminal Description
Hot (L1)
Neutral (N)
Protective Earth (PE)
1
2
11
Important Note: The AC power outlets cannot be used with the 24 VDC power supply.
AC power outlets can be installed to connect AC powered sc-sensors to the sc1000.
Refer to
Figure 16
and the following directions for proper power socket installation.
1. Remove power from the instrument.
2. Remove the probe module cover and remove the voltage barrier cover.
3. Remove the plug from the power socket housing opening. Feed the power socket
cable through the housing opening.
Wire Color Code for North America
3
4
5
10
8.
AC power connections
9.
Earth ground connection
10. Power outlet connection
11. Connector for expansion slot #1
12. Sensor connections
13. Relay card connection
Installation
客服:400 688 9809
Black
White
Green
6
7
8
9
电话:0431-85178766
19