Figure 21
Proper Wire Preparation and Insertion
1.
Strip ¼-inch of insulation.
Figure 22
Relay Card
1.
Terminal Block–Refer to
Terminal
Designation
1
Relay 1 (normally closed contacts)
2
3
Relay 2 (normally closed contacts)
4
5
Relay 3 (normally closed contacts)
6
7
Relay 4 (normally closed contacts)
8
网址:www.mamots.com
3. Properly prepare each wire and insert each wire
according to
Figure 22
connection is secure.
4. Reinstall the relay and probe module covers.
5. Refer to
Configuring New Components on page 54
1
2.
Seat insulation against connector with no bare wire exposed.
Table 8
for terminal assignments.
Table 8 Relay Card Terminal Assignments
and
Table
8. Tug gently after each insertion to ensure the
1
2
3
4
5
6
7
8
1
Relay 1–4
Maximum switching voltage:
250 VAC, 125 VDC
Nominal switching current:
250 VAC, 5 A
125 VAC, 5 A
125 VDC, 0.15 A
30 VDC, 5 A
Maximum switching capacity:
1500 VA 150 W
Installation
客服:400 688 9809
(Figure
21) into the terminal
for operation setup instructions.
2
电话:0431-85178766
25