Switching Measurement Result Displays - Konica Minolta BC-10 Plus Instruction Manual

Baking meter
Table of Contents

Advertisement

Switching Measurement Result Displays

It is possible to switch how the measurement results are displayed.
<Measurement>
10:00
05
05
3.55
BCU
61.1
L*
61.1
L*
7.2
a*
7.2
a*
21.3
b*
21.3
b*
<Average>
BCU statistics display
Avg. 05
AVERAGE
BCU
3.56
Avg
3.58
Max
3.53
Min
MENU/OK
MENU/OK
Calculation Methods
BCU (Baking Contrast Units)
"O" will be displayed when BCU value is less than zero (BCU ≤ 0).
Average of BCU
The averages of the measured tristimulus values X, Y, and Z of specimens are calculated
and the BCU value is calculated from the average Y value.
10:00
10:00
05
3.55
BCU
Average value of L*a*b*
10:00
10:00
Avg. 05
L* a* b* Avg.
AVERAGE
AVERAGE
61.1
L*
7.2
a*
21.3
b*
10:00
05
3.55
BCU
61.1
L*
7.2
a*
21.3
b*
Maximum value of L*a*b*
Minimum value of L*a*b*
10:00
Corresponding
data No.
Avg. 05
Avg. 05
Max 01
Min 03
AVERAGE
61.5
L*
L*
7.2
a*
a*
21.4
b*
b*
Average of L*a*b*
The averages of the measured tristimulus values X, Y, and Z of specimens are calculated
and the L*a*b* values are calculated from the average tristimulus values.
Maximum or minimum value of L*a*b*
The maximum or minimum of the measured tristimulus value Y is found and the L*a*b*
values are calculated using the X, Y, and Z values for the measurement corresponding to
the maximum or minimum Y value.
10:00
60.6
7.1
21.1
30