Avg. Meas.

Using the average measurement function makes it possible to obtain data that is even
more accurate.
If the sample color is uneven, the accuracy of the measurement can be improved by
randomly measuring locations within the sample area to calculate the average for the
sample, or by repeatedly measuring the same sample location to calculate the average
for that location.
Notes
After an average is determined for the instrument, because the average is
converted for each color space and displayed, the simple averages for
measurements of each count may not match.
1
Verify that the <Measurement> screen is
displayed.
If the measurement count does not read "00", pressing and
holding the AVERAGE button will clear the display.
2
Set the sample surface aperture on the sample to
measure.
3
Press the measuring button.
Beep
4
Measurement will be performed, and
the measurement results will be
displayed on the screen.
Continue performing measurements to
measure more than once.
To perform measurement again, press
MENU/OK and navigate to "Undo" to perform
the measurement again.
10:00
00
BCU
L*
a*
b*
10:00
Measurement
05
3.70
count
BCU
63.7
L*
7.3
a*
28.9
b*
5
While the measurement is displayed,
quickly press the AVERAGE button.
The average of the measurement results up
to that point will be displayed.
Press the measurement button again to
continue measuring.
6
Reset the average.
Press and hold the AVERAGE button for at least 1 second
to reset the measurement results.
Memo
It is also possible to reset the results by navigating to the
MENU screen and pressing
Reset.
Indicates that
Avg. 05
BCU
values are
average for fi ve
3.71
Avg
measurements
3.72
Max
3.70
Min
or
to select Avg.
29