Level; Customer Value; Activate Table - Endress+Hauser Levelflex FMP55 Operating Instructions Manual

Profibus pa guided wave radar
Hide thumbs Also See for Levelflex FMP55:
Table of Contents

Advertisement

Levelflex FMP55 PROFIBUS PA
Level (Manual)
Navigation
Prerequisite
Description
User entry
Level (Semiautomatic)
Navigation
Prerequisite
Description

Customer value

Navigation
Prerequisite
Description
User entry

Activate table

Navigation
Prerequisite
Description
Selection
Endress+Hauser
Setup → Advanced setup → Linearization → Level
• Linearization type (→  125) = Table
• Table mode (→  129) = Manual
Enter level value of the table point (value before linearization).
Signed floating-point number
Setup → Advanced setup → Linearization → Level
• Linearization type (→  125) = Table
• Table mode (→  129) = Semiautomatic
Displays measured level (value before linearization). This value is transmitted to the table.
Setup → Advanced setup → Linearization → Customer value
Linearization type (→  125) = Table
Enter linearized value for the table point.
Signed floating-point number
 Setup → Advanced setup → Linearization → Activate table
Linearization type (→  125) = Table
Activate (enable) or deactivate (disable) the linearization table.
• Disable
• Enable