Medium Group; Empty Calibration - Endress+Hauser Levelflex FMP55 Operating Instructions Manual

Profibus pa guided wave radar
Hide thumbs Also See for Levelflex FMP55:
Table of Contents

Advertisement

Operating menu

Medium group

Navigation
Prerequisite
Description
Selection
Additional information

Empty calibration

Navigation
Description
User entry
Factory setting
102
 Setup → Medium group
• For FMP51/FMP52/FMP54/FMP55: Operating mode (→  100) = Level
• Medium type = Liquid
Select medium group.
• Others
• Water based (DC >= 4)
This parameter roughly specifies the dielectric constant (DC) of the medium. For a more
detailed definition of the DC use the Medium property parameter.
The Medium group parameter presets the Medium property parameter as follows:
Medium group
Others
Water based (DC >= 4)
The Medium property parameter can be changed at a later point of time. However,
when doing so, the Medium group parameter retains its value. Only the Medium
property parameter is relevant for the signal evaluation.
The measuring range may be reduced for small dielectric constants. For details refer
to the Technical Information (TI) of the respective device.
 Setup → Empty calibr.
Distance process connection to min. level.
Depending on the probe
Depending on the probe
Levelflex FMP55 PROFIBUS PA
Medium property
Unknown
DC 4 ... 7
Endress+Hauser