Figure 2-18 shows the Terminal Module map for the HP E1419.
Figure 2-19 shows the Terminal Module map for the HP E1419.
Option A3F allows an HP E1419 to be connected to an HP E1586 Rack
Mount Terminal Panel. The option provides 4 SCSI plugs on a Terminal
Module to make connections to the Rack Mount Terminal Panel using 4
separately ordered SCSI cables. Option A3F Is shown in Figure 2-20.
There are two different cables available to connect the HP E1586 Rack
Mount Terminal Panel to the HP E1419 Option A3F. In both cases, four
cables are required if all 64-channels are needed. These cables do not come
with the HP E1419A Option A3F and must be ordered separately.
Standard Cable
This cable (HP E1588A) is a 16-channel twisted pair cable with an outer
shield. This cable is suitable for relatively short cable runs.
Custom Length Cable
This cable (HP Z2220A Option 050) is available in custom lengths. It is a
16-channel twisted pair cable with each twisted pair individually shielded to
provide better quality shielding for longer cable runs.
HF Common Mode Filters
Optional High Frequency Common Mode Filters are on the HP E1586 Rack
Mount Terminal Panel's input channels (HP E1586 Option 001, RF Filters).
They filter out AC common mode signals present in the cable that connects
between the terminal panel and the device under test. The filters are useful
for filtering out small common mode signals below 5 Vp-p. To order these
filters, order HP E1586 Option 001.