282 HP E1419 Command Reference
is used to compensate all subsequent thermocouple measurements until the register
is overwritten by another reference measurement or by specifying a constant
reference temperature with the SENSE:REF:TEMP command. If used, the reference
junction channel must be scanned before any thermocouple channels. Use the
SENSE:REF:CHANNELS command to place the reference measuring channel into
the scan list ahead of the thermocouple measuring channels.
Parameter
Parameter
Name
Type
type
discrete (string)
sub_type
numeric (float32)
numeric (float32)
range
numeric (float32)
ch_list
channel list (string)
See "Linking Input Channels to EU Conversion" on page 3-59 for more
information.
The <range> parameter: The HP E1419 has five ranges: .0625VDC, .25VDC,
1VDC, 4VDC, and 16VDC. To select a range, simply specify the range value (for
example, 4 selects the 4VDC range). If you specify a value larger than one of the
first four ranges, the HP E1419 selects the next higher range (for example, 4.1
selects the 16VDC range). Specifying a value larger than 16 causes an error.
Specifying 0 selects the lowest range (.0625VDC). Specifying AUTO selects auto
range. The default range (no range parameter specified) is auto range.
If you are using amplifier SCPs, you should set them first and keep their settings
in mind when specifying a range setting. For instance, if your expected signal
voltage is to be approximately .1VDC and the amplifier SCP for that channel has
a gain of 8, you must set <range> no lower than 1VDC or an input out-of-range
condition will exist.
The <type> parameter specifies the sensor type that will be used to determine the
temperature of the isothermal reference panel. <type> CUSTom is pre-defined as
Type E with 0°C reference junction temp and is not re-defineable.
For <type> THERmistor, the <sub_type> parameter may be specified in ohms or
kohm.
The *CAL? command calibrates resistance channels based on Current Source
SCP and Sense Amplifier SCP setup at the time of execution. If SCP settings are
changed, those channels are no longer calibrated. *CAL? must be executed again.
Range of
Values
THERmistor | RTD | CUSTom
for THER use 5000
for RTD use 85 | 92
for CUSTom use 1
see comments
100 - 163
Default
Units
none
Ohm
none
none
VDC
none
Chapter 6