[Sense:]Function:totalize; [Sense:]Function:voltage[:Dc]; [Sense:]Function:totalize (@<Ch_List>); [Sense:]Function:voltage[:Dc] [<Range>,](@<Ch_List>) - HP E1419A User Manual

Measurement and control module
Table of Contents

Advertisement

[SENSe]
Usage

[SENSe:]FUNCtion:TOTalize

Parameters
Comments
Usage

[SENSe:]FUNCtion:VOLTage[:DC]

280 HP E1419 Command Reference
Link two channels to the K type thermocouple temperature conversion
SENS:FUNC:TEMP TCOUPLE,K,(@101,102)
Link channel 0 to measure reference temperature using 5K thermistor
SENS:REF THER,5000,(@100)
[SENSe:]FUNCtion:TOTalize <ch_list> sets the SENSe function to TOTalize for
channels in <ch_list>.
Parameter
Parameter
Name
Type
ch_list
string
The totalize function counts rising edges of digital transitions at
Frequency/Totalize SCP channels. The counter is 24 bits wide and can count up to
16,777,215.
The SENS:TOT:RESET:MODE command controls which events will reset the
counter.
If the channels specified are not on a Frequency/Totalize SCP, an error will be
generated.
Related Commands: SENS:TOT:RESET:MODE, INPUT:POLARITY
*RST Condition: SENS:FUNC:COND and INP:POL NORM for all digital SCP
channels.
SENS:FUNC:TOT (@148)
[SENSe:]FUNCtion:VOLTage[:DC] [<range>,](@<ch_list>) links the specified
channels to return DC voltage.
Range of
Values
132 - 163
channel 48 is a totalizer
Default
Units
none
Chapter 6