Subsystem Syntax

OUTPut:CURRent:AMPLitude

NOTE
Chapter 6
The OUTPut subsystem is involved in programming source SCPs as well as
controlling the state of VXIbus TTLTRG lines 0 through 7.

OUTPut

:CURRent
:AMPLitude <amplitude>,(@<ch_list>)
:AMPLitude? (@<channel>)
[:STATe] 1 | 0 | ON | OFF,(@<ch_list>)
[:STATe]? (@<channel>)
:POLarity NORMal | INVerted,(@<ch_list>)
:POLarity? (@<channel>)
:SHUNt 1 | 0 | ON | OFF,(@<ch_list>)
:SHUNt? (@<channel>)
:TTLTrg
:SOURce TRIGger | FTRigger | SCPlugon | LIMit
:SOURce?
:TTLTrg<n>
[:STATe] 1 | 0 | ON | OFF
[:STATe]?
:TYPE PASSive | ACTive,(@<ch_list>)
:TYPE? (@<channel>)
:VOLTage
:AMPLitude <amplitude>,(@<ch_list>)
:AMPLitude? (@<channel>)
OUTPut:CURRent:AMPLitude <amplitude>,(@<ch_list>) sets the HP E1505
Current Source SCP channels specified by ch_list to either 488 µA, or 30 µA. This
current is typically used for four-wire resistance and resistance temperature
measurements.
This command does not set current amplitude on SCPs like the HP E1532 Current
Output SCP.