Output:shunt; Output:shunt? (@<Channel>) - HP E1419A User Manual

Measurement and control module
Table of Contents

Advertisement

OUTPut:SHUNt

Parameters
Comments
Usage
OUTPut:SHUNt?
Parameters
Comments
Chapter 6
Returned Value: returns one of NORM or INV. The type is string.
OUTPut:SHUNt <enable>,(@<ch_list>) adds shunt resistance to one leg of bridge
on Strain Bridge Completion SCPs. This can be used for diagnostic purposes and
characterization of bridge response.
Parameter
Parameter
Name
Type
enable
boolean (uint16)
ch_list
channel list (string)
If ch_list specifies a non strain SCP, a 3007 "Invalid signal conditioning plug-on"
error is generated.
When Accepted: Not while INITiated
Related Commands: [SENSe:]FUNCtion:STRain..., [SENSe:]STRain...
*RST Condition: OUTP:SHUNT 0 on all Strain SCP channels
OUTP:SHUNT 1,(@148:151)
OUTPut:SHUNt? (@<channel>) returns the status of the shunt resistance on the
specified Strain SCP channel.
Parameter
Parameter
Name
Type
channel
channel list (string)
channel must specify a single channel only.
If channel specifies a non strain SCP, a 3007 "Invalid signal conditioning
plug-on" error is generated.
Returned Value: Returns 1 or 0. The C-SCPI type is uint16.
Related Commands: OUTP:SHUNT
Range of
Values
0 | 1 | ON | OFF
132 - 163
add shunt resistance at channels 48 - 51
Range of
Values
132 - 163
HP E1419 Command Reference 253
OUTPut
Default
Units
none
none
Default
Units
none

Advertisement

Table of Contents
loading

Table of Contents