Status:operation:enable; Status:operation:enable <Enable_Mask - HP E1419A User Manual

Measurement and control module
Table of Contents

Advertisement

Usage

STATus:OPERation:ENABle

Parameters
Comments
Usage
Chapter 6
Returned Value: Decimal weighted sum of all set bits. The C-SCPI type is uint16.
Related Commands: *CAL?, CAL:ZERO, INITiate[:IMMediate],
STAT:OPER:EVENT?, STAT:OPER:ENABLE, STAT:OPER:ENABLE?
*RST Condition: No Change
STATUS:OPERATION:CONDITION?
STATus:OPERation:ENABle <enable_mask> sets bits in the Enable register that
will enable corresponding bits from the Event register to set the Operation summary
bit.
Parameter
Parameter
Name
Type
enable_mask
numeric (uint16)
Enable_mask may be sent as decimal, hex (#H), octal (#Q), or binary (#B).
VXI Interrupts: When Operation Status Group bits 4, 8, 9, 10, or 11 are enabled,
VXI card interrupts will occur as follows:
When the event corresponding to bit 4 occurs and then is cleared, the card
will generate a VXI interrupt. When the event corresponding to bit 8, 9, 10,
or 11 occurs, the card will generate a VXI interrupt.
NOTE: In C-SCPI, the C-SCPI overlap mode must be on for VXIbus
interrupts to occur.
Related Commands: *STB?, SPOLL, STAT:OPER:COND?,
STAT:OPER:EVENT?, STAT:OPER:ENABLE?
Cleared By: STAT:PRESet and power-on.
*RST Condition: No change
STAT:OPER:ENABLE 1
Enter statement will return value from
condition register
Range of
Values
0-32767
Set bit 0 in the Operation Enable register
HP E1419 Command Reference 299
STATus
Default
Units
none