[Sense:]Strain:unstrained; [Sense:]Totalize:reset:mode; [Sense:]Strain:unstrained? (@<Channel>); [Sense:]Totalize:reset:mode Init | Trigger,(@<Ch_List>) - HP E1419A User Manual

Measurement and control module
Table of Contents

Advertisement

[SENSe]

[SENSe:]STRain:UNSTrained?

Parameters
Comments
Usage

[SENSe:]TOTalize:RESet:MODE

Parameters
Comments
Usage
288 HP E1419 Command Reference
[SENSe:]STRain:UNSTrained? (@<channel>) returns the unstrained voltage value
currently set for the sense channel specified by <channel>. This command does not
make a measurement.
Parameter
Parameter
Name
Type
channel
channel list (string)
Returned Value: Numeric value of unstrained voltage. The C-SCPI type is flt32.
<channel> must specify a single channel only.
Related Commands: STRAIN:UNST
STRAIN:UNST? (@107)
enter statement here
[SENSe:]TOTalize:RESet:MODE <select>,<ch_list> sets the mode for resetting
totalizer channels in <ch_list>.
Parameter
Parameter
Name
Type
select
discrete (string)
ch_list
string
In the INIT mode the total is reset only when the INITiate command is executed.
In the TRIGger mode the total is reset every time a new scan is triggered.
If the channels specified are not on a Frequency/Totalize SCP, an error will be
generated.
Related Commands: SENS:FUNC:TOT, INPUT:POLARITY
*RST Condition: SENS:TOT:RESET:MODE INIT
SENS:TOT:RESET:MODE TRIG,(@148)
Range of
Values
100 - 163
query unstrained voltage for channel 7
returns the unstrained voltage set by
STR:UNST
Range of
Values
INIT | TRIGger
132 - 163
totalizer at channel 48 resets at each trigger
event
Default
Units
none
Default
Units
seconds
none
Chapter 6

Advertisement

Table of Contents
loading

Table of Contents