Page 1
sales@artisantg.com artisantg.com (217) 352-9330 | Click HERE Find the Keysight / Agilent E1509A at our website:...
Page 2
HP E1509 User’s Manual The HP E1503 manual also applies to HP E1413Bs as HP E1413 Option 19. Enclosed is the User’s Manual for the HP E1509 Signal Conditioning Plug-on. Insert this manual in your HP E1413/E1313 manual behind the “Signal Conditioning Plug-ons”...
Chapters 1 and 2 of your HP E1413/E1313 manual. Identifying the Plug-on You’ll find the HP part number on the connector side of the SCP to the left of the serial number bar code. For the HP E1509, the part number is :...
The connections are shown in Figures 2 and 3. • For HP E1413B and below Terminal Modules, use stickers for HP E1413 Option 19 SCPs. The connections are shown in Figure 4. Figure 1 HP E1509 C-Size Terminal Module Connections 4 HP E1509 Fixed Filter+Amplifier SCP...
Page 6
Figure 2 HP E1509 B-size Terminal Module Connections (Ch 00-31) Figure 3 HP E1509 B-size Terminal Module Connections (Ch 32-63) Figure 4 HP E1413 Option 19 Terminal Module Connections Connecting To The Terminal Module HP E1509 Fixed Filter+Amplifier SCP 5...
HP E1413/E1313 manual. This section will relate those commands to the parameter values which are specific to this Plug-on. Checking the ID To verify the SCP type(s) installed on the HP E1413/E1313 use the SYSTem:CTYPe? (@<channel>) command. of the SCP •...
Page 8
2 enter statement here Querying the While the HP E1509’s amplifiers have fixed gain, the channel gain can be queried. The response to this query will always be 64. To query any channel Channel Gain to determine its gain setting use the INPut:GAIN? (@<channel>) command.
Register Based Programming The register-based commands shown here are covered in Appendix D of the HP E1413/E1313 manual. You should read that section first to become familiar with accessing registers and executing Register-Based Commands. This section will relate those commands to the parameter values which are specific to this Plug-on.
Page 10
Bit 7 Bit 6 Bit 5 Bit 4 Bit 3 Bit 2 Bit 1 Bit 0 SCP 7 SCP 6 SCP 5 SCP 4 SCP 3 SCP 2 SCP 1 SCP 0 Register Based Programming HP E1509 Fixed Filter+Amplifier SCP 9...
Specifications These specifications for the HP E1509 reflect the combined performance of the HP E1413/E1313 and the HP E1509 Signal Conditioning Plug-on. These specifications are not to be added to those presented in the HP E1413/E1313 User’s Manual. General Specifications Measurement ranges ±3.9mV to ±.25V FS...
Page 12
100 to 200 °C 200 to 600 °C 600 to 1000 °C Type R A/D Filter 1.40°C 0.75°C 0.45°C 0.30°C 1.40°C 0.70°C 0.40°C 0.30°C * [SENSe:]FILTer[:LPASs][:STATe] ON (max scan rate - 100 rdgs/sec/channel) Specifications HP E1509 Fixed Filter+Amplifier SCP 11...
Page 13
(simplified specifications, see temperature accuracy graphs in HP E1413/E1313 Measurement accuracy manual for details) Temperature (cont.) Thermocouples (cont.) 0 to 100 °C 100 to 200 °C 200 to 800 °C 800 to 1750 °C Type S A/D Filter 2.85°C 1.35°C 0.70°C...