Trigger:count; Trigger:count <Trig_Count - HP E1419A User Manual

Measurement and control module
Table of Contents

Advertisement

TRIGger

TRIGger:COUNt

Parameters
Comments
Usage

TRIGger:COUNt?

Comments
Usage
312 HP E1419 Command Reference
TRIGger:COUNt <trig_count> sets the number of times the module can be
triggered before it returns to the Trigger Idle State. The default count is 0 (same as
INF) so accepts continuous triggers. See Figure 6-6
Parameter
Parameter
Name
Type
Trig_count
numeric (uint16)
(string)
When trig_count is set to 0 or INF, the trigger counter is disabled. Once INITiated
the module will return to the Waiting For Trigger State after each trigger event.
The ABORT (preferred) and *RST commands will return the module to the
Trigger Idle State. ABORT is preferred since *RST also returns other module
configurations to their default settings.
The default count is 0
Related Commands: TRIG:COUNT?
*RST Condition: TRIG:COUNT 0
TRIG:COUNT 10
TRIG:COUNT 0
TRIGger:COUNt? returns the currently set trigger count.
If TRIG:COUNT? returns 0, the trigger counter is disabled and the module will
accept an unlimited number of trigger events.
Returned Value: Numeric 0 through 65,535. The C-SCPI type is int32.
Related Commands: TRIG:COUNT
*RST Condition: TRIG:COUNT? returns 0
TRIG:COUNT?
enter statement
Range of
Values
0 to 65535 | INF
Set the module to make 10 passes all
enabled algorithms.
Set the module to accept unlimited triggers
(the default)
Query for trigger count setting
Returns the TRIG:COUNT setting
Default
Units
none