HP E1419A User Manual page 324

Measurement and control module
Table of Contents

Advertisement

NOTE
Chapter 6
Executing *TST? returns the module to its *RST state. *RST causes the FIFO data
format to return to its default of ASC,7. If you want to read the FIFO for *TST?
diagnostic information and you want that data in other than the ASCII,7 format, be
certain to set the data FIFO format to the desired format (FORMAT command) after
completion of *TST? but before executing a SENSE:DATA:FIFO: query command.
The C-SCPI type for this returned value is int16.
Following *TST?, the module is placed in the *RST state. This returns many of
the module's programmed states to their defaults. See page 3-53 for a list of the
module's default states.
*TST? performs the following tests on the HP E1419 and installed Signal
Conditioning Plug-ons:
DIGITAL TESTS:
Test#
1-3:
Writes and reads patterns to registers via A16 & A24
4-5:
Checks FIFO and CVT
6:
Checks measurement complete (Measuring) status bit
7:
Checks operation of FIFO half and FIFO full IRQ generation
8-9:
Checks trigger operation
ANALOG FRONT END DIGITAL TESTS:
Test#
20:
Checks that SCP ID makes sense
30-32:
Checks relay driver and fet mux interface with EU CPU
33,71:
Checks opening of all relays on power down or input overvoltage
34-37:
Check fet mux interface with A/D digital
ANALOG TESTS:
Test#
40-42:
Checks internal voltage reference
ANALOG TESTS: (continued)
Test#
43-44:
Checks zero of A/D, internal cal source and relay drives
45-46:
Checks fine offset calibration DAC
47-48:
Checks coarse offset calibration DAC
49:
Checks internal + and -15V supplies