CALibration
Command
Sequence

CALibration:TARE:RESet

Command
Sequence

CALibration:TARE?

218 HP E1419 Command Reference
When Accepted: Not while INITiated
Related Commands: CAL:TARE?, CAL:STOR TARE
*RST Condition: Channel offsets are not affected by *RST.
CAL:TARE < ch_list >
CAL:TARE?
CAL:STORE TARE
CALibration:TARE:RESet resets the tare calibration constants to zero for all
64 channels. Executing CAL:TARE:RES affects the tare cal constants in RAM only.
To reset the tare cal constants in Flash Memory, execute CAL:TARE:RES and then
execute CAL:STORE TARE.
CAL:TARE:RESET
CAL:STORE TARE
CALibration:TARE? Returns a value to indicate the success of the last CAL:TARE
operation. CAL:TARE? returns the value only after the CAL:TARE operation is
complete.
Returned Value:
Value
0
-1
-2
The C-SCPI type for this returned value is int16.
Executing CAL:TARE sets the Calibrating bit (bit 0) in Operation Status Group.
Executing CAL:TARE? resets the bit.
to correct channel offsets
to return the success flag from the
CAL:TARE operation
Optional depending on necessity of long
term storage
to reset channel offsets
Optional if necessary to reset tare cal
constants in Flash Memory.
Meaning
Cal OK
Cal Error
Query the Error Queue (SYST:ERR?)
See Error Messages in Appendix B.
No results available
Further Action
None
Also run *TST?
Perform CAL:TARE
before CAL:TARE?
Chapter 6