364 Error Messages
2016
'Byte count is not a multiple of two'.
3000
'Illegal while initiated'. Operation must be performed
before INIT or INIT:CONT ON.
3004
'Illegal command. CAL:CONF not sent'. Incorrect
sequence of calibration commands. Send
CAL:CONF:VOLT command before CAL:VAL:VOLT
and send CAL:CONF:RES command before
CAL:VAL:RES
3005
'Illegal command. Send CAL:VAL:RES'. The only
command accepted after a CAL:CONF:RES is a
CAL:VAL:RES command.
3006
'Illegal command. Send CAL:VAL:VOLT'. The only
command accepted after a CAL:CONF:VOLT is a
CAL:VAL:VOLT command.
3007
'Invalid signal conditioning module'. The command sent
to an SCP was illegal for its type.
3008
'Too few channels in scan list'. A Scan List must contain
at least two channels.
3012
'Trigger too fast'. Scan list not completed before another
trigger event occurs.
3015
'Channel modifier not permitted here'.
3019
'TRIG:TIM interval too small for SAMP:TIM interval
and scan list size'. TRIG:TIM interval must allow for
completion of entire scan list at currently set SAMP:TIM
interval. See TRIG:TIM in Chapter 5, the Command
Reference
3020
'Input overvoltage'. Calibration relays opened (if
JM2202 not cut) to protect module inputs, and
Questionable Data Status bit 11 set. Execute *RST to
close relays and/or reset status bit.
3021
'FIFO overflow'. Lets you know that the FIFO buffer has
filled and that one or more readings have been lost.
Usually caused by algorithm values stored in FIFO faster
than FIFO was read.
3026
'Calibration failed'.
3027
'Unable to map A24 VXI memory'.
Appendix B