Memory Errors

The following may occur if the cable connecting the instrument to the PC is disconnected.
• It may become impossible to save fi les to the instrument.
• Files saved on the instrument may become unreadable.
Data recovery can be performed if such memory errors occur.
Notes
After a memory error occurs, the data will be recovered. Although
measurement is possible, the measurement data cannot be saved to the
instrument.
Recover may take several minutes. During the recovery process, all
operations will be prohibited.

When not performing recovery

Notes
Only measurement will be possible if recovery is not performed. When an
error occurs, measurement data cannot be saved even if "Save to
Instrument" is set. An error message will be displayed every time the
power is turned on until system recovery is performed.
1
Select "No" and confi rm.
MENU/OK
or
2
Review the message and press "OK" to confi rm.
MENU/OK
The display will return to the measurement screen.

Performing Recovery

1
Select "Yes" and confi rm.
2
Review the message and press "OK" to confi rm.
MENU/OK
The power will turn OFF automatically.
10:00
Memory Error
Memory recovery
needed?
Time
required:
approx.
10 min.
3
Press
No
Yes
A message is displayed after the power is turned on.
10:00
Measurement
can be performed,
but data
cannot be
saved to the
instrument.
OK
MENU/OK
or
POWER
again to turn the power ON.
76
10:00
Memory Error
Memory recovery
needed?
Time
required:
approx.
10 min.
No
Yes
10:00
Memory
recovery
will be
performed.
Turn the
power off
and back on.
OK
10:00
Perform
memory
recovery
after
connecting
to a power supply.
No
Yes