Fresenius Medical Care 2008K Troubleshooting Manual page 89

Hemodialysis system
Hide thumbs Also See for 2008K:
Table of Contents

Advertisement

Table 5 – Noted Valve Checks
To check valve #43
To check valves #39,
#29 and #43
To check valves #30
and #26
To check valves #31
and #34
To check valves #32
and #33
Table 6 (Valve #41, #27)
VALVE#
PIN#
from the front of the
41
32
Table 7 (Valves #32 and #33)
VALVE#
PIN#
32
37
front of machine
33
38
from the front of
See
Table 8
(page 88)
See
Table 14
(page 91)
See
Table 13
(page 90)
See
Table 10
(page 89)
See
Table 7
(page 87)
P2 LOCATION
Top row, 10 pins
machine
P2 LOCATION
Between 40 AND 100
row in this table).
Bottom row, 7
pins from the
Less than 39
Greater than 101
2) Bad distribution board
Between 40
and 100
Top row, 7 pins
Less than 39
the machine
Greater than 101
2) Bad distribution board
Fresenius 2008K Troubleshooting Guide
P/N 507298 Rev. B (Draft 1)
To check valves #35
and #38
To check valves #36
and #37
To check valve #25
To check valve #41
(#27)
Resistance Reading (see column below)
Between 40 AND 100
Less than 39
: Bad actuator board
Greater than 101
: Two possibilities: 1) Bad actuator
cable; 2) Bad distribution board
Resistance Reading (see column below):
: Check valve #33 (pin #38, next
: Bad actuator board.
: Two possibilities: 1) Bad actuator board;
AND 100
: If BOTH valves are between 40
the actuator board is bad
: Bad actuator board.
: Two possibilities: 1) Bad actuator cable;
Troubleshooting Valve Errors
See
Table 9
(page 88)
See
Table 11
(page 89)
See
Table 12
(page 90)
See
Table 6
(page 87)
: Bad actuator board.
Page 87