TMP- 1.4.2 CHECK PDIAL (DIALYSATE PRESSURE)
a) Clamp the four way's syringe tubing to hold pressure!
b) Call debug screen 0. If valve #25's dot IS blue (see Figure
right) then valve #25 is open. Continue to part c. If the dot is
NOT blue increase VEN (screen 1). If unable to make the dot
blue remove the tubing from the P
number
TMP- 6.0.0
c) Monitor the PDial data window (see Figure right). Two scenarios:
1) PDial AGREES with the external meter within +/- 25 mmHg: See procedure number
TMP- 1.4.5
(page 184).
2) PDial does NOT agree with the external meter within +/- 25 mmHg: See procedure number
TMP- 6.0.0
(page 187).
TMP- 1.4.3 ISOLATE SECONDARY CIRCUIT
a) Be CAREFUL HERE!
#43's output nozzle (towards the FRONT of the machine).
hydraulic circuit isolation area.
b) Can you create approximately -250 mmHg AND
Yes
See procedure number
No
a) Open the shunt door (valve #24 and #25 close).
isolation area.
b) Pull on the syringe then clamp the syringe tubing. Can you create approximately -250
mmHg AND hold it (+/-20 mmHg per minute) on the external meter now?
Yes
No
port then see procedure
ven
(page 187).
Referring to
Figure 69
TMP- 1.4.7
There is a large negative pressure leak in the secondary side hydraulics.
Remove the clamps from valve #43 then see procedure number
(page 189).
a) Remove the clamps from valve #43!
b) Check the four way assembly connections for a leak. If SURE that the
assembly is NOT leaking see procedure number
Figure 69 – Clamping Valve #43
Fresenius 2008K Troubleshooting Guide
P/N 507298 Rev. B (Draft 1)
(page 183),
DOUBLE clamp the tubing at valve
Figure 73 B
(
+/- 20 mmHg per minute) now?
hold it
(page 184).
Figure 73 C
Section 9 –TMP Problems
(page 197) illustrates the
(page 197) illustrates the
TMP- 8.1.0
TMP- 8.1.0
(page 189).