Fresenius Medical Care 2008K Troubleshooting Manual page 182

Hemodialysis system
Hide thumbs Also See for 2008K:
Table of Contents

Advertisement

IMPORTANT!
Yes
a)
position labeled "AIR-SEN". Allow 30 seconds then ASSURE FILACT returns to = 0.
b) See procedure number
No
a) Not likely! ASSURE that distribution board position "AIR-SEN" is EMPTY before
continuing! If not, repeat procedure number
b) After 30 seconds, If (and ONLY if) FILACT still = 0, Filling Program is malfunctioning.
Proceed to page 17, procedure number
TMP- 1.0.10 DETERMINE TMP PROBLEM
a) Call the Home screen. Assure Dialysate flow is set to 500 ml/min.
b) Referring to
Figure 67
upper and lower alarm Limits (at 520 and 0) and an 'actual' TMP bar which is either red (with a TMP
alarm) or green (no TMP alarm).
c) Press and release the 'RESET' key then immediately press and hold it for three seconds. If the
"Adjusting TMP" message is present allow it to expire. Ignore temporary Flow Errors during reset! If
a TMP alarm occurs attempt the RESET sequence up to twice more then continue.
d) Watch the (green or red) 'actual' TMP bar. TMP is STABLE if (and ONLY if) it REMAINS between the 0
and 520 limits AND varies less than +/- 40 mmHg in two minutes. Proceed accordingly (three
scenarios):
1. TMP is UNSTABLE if the 'actual' TMP bar rises ( ), greater than +40, towards the 520
Limit OR exceeds the 520 Limit causing a "TMP Is High" alarm: See procedure number
TMP- 1.0.15 (page 181).
2. TMP is UNSTABLE and the 'actual' TMP bar falls ( ), greater than -40, towards the lower
0 Limit OR falls below the 0 Limit causing a "TMP Is LOW" alarm: See procedure number
TMP- 1.0.15 (page 181).
3. TMP is STABLE AND the 'actual' TMP bar is REMAINING between the 520 and 0 Limits: Is
the 'actual' TMP bar green?
Yes
See procedure number
No
See procedure number
CAREFULLY return the air sensor's connector to distribution board
TMP- 1.0.10
(page 180), the TMP bar graph is populated with three horizontal bars; yellow
Figure 67 – TMP Bar Graph
TMP- 1.3.0
TMP- 1.0.11
Fresenius 2008K Troubleshooting Guide
P/N 507298 Rev. B (Draft 1)
(page 180).
TMP- 1.0.9
(page 179).
F-
1.0.10.
(page 182).
(page 181).
Section 9 –TMP Problems
Page 180

Advertisement

Table of Contents
loading