Fresenius Medical Care 2008K Troubleshooting Manual page 75

Hemodialysis system
Hide thumbs Also See for 2008K:
Table of Contents

Advertisement

CLEAN- 2.0.4 ISOLATE VALVE #39 / ACTUATOR BOARD
a) IMPORTANT! Allow Rinse program to continue for ten minutes.
b) Referring to
Figure 102
chamber. Touch valve #39's (black) solenoid. Is it warm (i.e. energized)?
Yes
The actuator board is okay. Valve #39 is bad.
No
NOTE that VALVE #39 (ONLY) will be checked and proceed to page 78,
VALVES.
CLEAN- 2.0.5 DEAP ADJUSTED?
As NOTED above was it necessary to adjust DEAP?
Yes
See procedure number
No
Three possibilities (see
deaeration motor (try brushes); 3) Bad loading pressure valve #65. IMPORTANT! After
repair see procedure number
CLEAN- 2.0.6 VERIFY PROBLEM
Place the machine into Heat Disinfect. Three scenarios:
1) Heat Disinfect times out (completes) successfully: See procedure number
2) A Flow Error returns AND the deaeration motor STOPS rotating: Two possibilities (see
page 71): 1) Bad deaeration pump head; 2) Bad deaeration motor (try brushes); IMPORTANT! After
repair see procedure number
3) A Flow Error returns AND the deaeration motor does NOT STOP rotating: Check loading
pressure. Is it peaking to 18 psi or greater?
Yes
a) Return the red concentrate connector to its rinse port.
IMPORTANT!
b)
c) See procedure number
No
Three possibilities (see
deaeration motor (try brushes); 3) Bad loading pressure valve #65. IMPORTANT! After
repair see procedure number
CLEAN- 2.0.7 CHECK DEAERATION AND LOADING PRESSURE
Check DEAERATION AND LOADING PRESSURE per Preventative Maintenance Procedures and calibrate
if necessary. NOTE: The Calibration screen's [Pump Rate] data box MUST be between 180 and 215
CLEAN- 3.0.0 CHECK VALVE CIRCUITS (VERR)
Call debug screen 1. Is VERR (lower right corner) greater than 5?
Yes
See procedure number
No
See procedure number CLEAN- 4.0.0 (page 74).
CLEAN- 3.0.1 CHECK VERR IN DIALYSIS PROGRAM
Any valve circuit can cause valve errors. These procedures isolate them:
a) Turn the machine off then back on (to reset VERR).
b) Return the concentrate connectors to concentrate and place the machine into Dialysis program
(remember to press the 'CONFIRM' key).
(page 246), valve #39 is mounted to the bottom of the hydro block or hydro
CLEAN- 2.0.6
Figure
32, page 71): 1) Bad deaeration pump head; 2) Bad
CLEAN- 2.0.7
CLEAN- 2.0.7
(page 73).
If the screen reads "Interrupted
CLEAN- 3.0.0
Figure
32, page 71): 1) Bad deaeration pump head; 2) Bad
CLEAN- 2.0.7
CLEAN- 3.0.1
(page 73).
Fresenius 2008K Troubleshooting Guide
P/N 507298 Rev. B (Draft 1)
Section 3 – Flow Errors In Cleaning Programs
(page 73).
(page 73).
"
restart Heat Disinfect.
(page 73).
(page 73).
TROUBLESHOOTING
CLEAN- 2.0.7
(page 73).
Figure
32,
Page 73