1. TMP is UNSTABLE if the 'actual' TMP bar rises ( ), greater than +40, toward the 520 Limit
OR exceeds the 520 Limit causing a "TMP Is High" alarm: See procedure number
(page 32).
2. TMP is UNSTABLE if the 'actual' TMP bar falls ( ), greater than -40, toward the lower 0
Limit OR falls below the 0 Limit causing a "TMP Is LOW" alarm: See procedure number
F- 6.6.0
(page 32).
3. TMP is STABLE: See procedure number
F- 6.6.0 TROUBLESHOOT FLOW ERROR (UNSTABLE TMP)
NOTE this procedure and page number. It may be necessary to perform all procedures below:
NOTE
a)
nine leak tests will be performed: 1) IOS; 2) Balancing Chamber Valves #31 through #38.
See
SECTION 18A – AUTOMATED VALVE LEAK TESTS
b) If a leak is not located, perform
DIAPHRAGM
(page 238).
c) If a diaphragm leak is not located, return to Dialysis program. If the Flow Error reoccurs see procedure
number
F- 1.0.2
(page 14).
F- 6.7.0 TROUBLESHOOT FLOW ERROR (STABLE TMP)
NOTE this procedure and page number. It may be necessary to perform all procedures below:
a) Possible torn balancing chamber diaphragm. Proceed to
BALANCING CHAMBER DIAPHRAGM
NOTE
b)
nine leak tests will be performed: 1) IOS; 2) Balancing Chamber Valves #31 through #38.
See
SECTION 18A – AUTOMATED VALVE LEAK TESTS
c) Return to Dialysis program (remember to press the 'CONFIRM
(constant or intermittent) return to procedure number
F- 7.0.0 DETERMINE DIASAFE
a) Assure the dialyzer quick connectors are in the shunt with the door closed.
b) Re-check loading pressure. If (and ONLY if) it is NOT 'okay' return to procedure number
18).
c) Call debug screen 0. Referring
Two scenarios:
1. NOT equipped with DiaSafe
been removed but Test Valve #91 remains installed. The knob MUST be
pointing to 'DIALYZE' (see Figure, right) otherwise flow is blocked at the valve.
Proceed to
F- 7.0.2
2. Equipped with a DiaSafe
®
DiaSafe
filter for kinks. If equipped with Test Valve #91, assure the knob is
pointing to 'DIALYZE' (see Figure, right) otherwise flow is blocked at the valve.
Proceed to
F- 7.0.2
F- 6.7.0
SECTION 19 - TESTING FOR A LEAKING BALANCING CHAMBER
(page 238).
®
OPTIONS
Figure 13
(page 33), how to proceed depends on DiaSafe
®
filter: In rare cases the DiaSafe
(page 33).
®
Filter: CAREFULLY check the tubing to and from the
(page 33).
Fresenius 2008K Troubleshooting Guide
P/N 507298 Rev. B (Draft 1)
Section 1 – Flow Errors In Dialysis Program
(page 32).
(page 230).
SECTION 19 - TESTING FOR A LEAKING
(page 230).
'
key). Call screen 0. If Flow Error = 1
F- 1.0.2
(page 14).
®
filter may have
F- 6.6.0
F- 2.0.4
(page
®
Filter options.
Page 32