TMP- 8.1.0 TROUBLESHOOT TMP
NOTE this procedure and page number in case you need to return here. Proceed until a problem is found:
a) Perform
SECTION 12 - INDUCED POSITIVE PRESSURE TESTS
located continue to part b. If a leak is located this is the problem!
b) Perform
SECTION 11 - INDUCED AIR LEAK TESTS
part c. If a leak is located this is the problem!
NOTE
c)
nine leak tests will be performed: 1) IOS; 2) Balancing Chamber Valves #31 through #38. See
SECTION 18A – AUTOMATED VALVE LEAK TESTS
d) Assuming all procedures have been performed correctly the Troubleshooting Guide is unable to
determine an immediate problem. If the TMP problem continues replace the DiaSafe
per procedure. If the TMP problem continues the dialysate pressure transducer (#9) may be bad
intermittent (see
Figure
TMP- 8.2.0 CALIBRATE DEAERATION PRESSURE
Calibrate DEAERATION AND LOADING PRESSURE per procedure. Is the calibration successful?
Yes
Return to Dialysis program (remember to press the 'CONFIRM' key). If the TMP problem
continues return to procedure number
No
Proceed to page 211,
70, page 185).
TMP- 1.0.6
SECTION 13 - DEAERATION
Fresenius 2008K Troubleshooting Guide
P/N 507298 Rev. B (Draft 1)
(page 208). If no external leaks are
(page 202). If no air leaks are located continue to
(page 230).
(page 178).
PROBLEMS.
Section 9 –TMP Problems
®