Fresenius Medical Care 2008K Troubleshooting Manual page 131

Hemodialysis system
Hide thumbs Also See for 2008K:
Table of Contents

Advertisement

Yes
Close the shunt door and see procedure number
No
Proceed to page 14,
Table 21 – Loading Pressure
®
DiaSafe
filter?
Yes
No
CO- 1.0.2 ACETATE OR BICARBONATE SELECTED?
This procedure determines if ACETATE or BICARBONATE concentrate is selected:
ACETATE is VERY rarely (if ever) used. Only one concentrate (a 1:34 ACETATE) is used and the
bicarbonate (blue) concentrate connector is left in its rinse port. It may have been selected
accidently.
If BICARBONATE is selected BOTH the red and blue concentrate connectors are plugged into
concentrate. In MOST cases BICARBONATE is selected:
a) Press the 'Escape' key to call the Home screen.
b) Select the 'Dialysate' tab, at the bottom of the screen, to call the
'CONFIRM Concentrate' screen (see
c) Is the screen's [Bicarbonate] window = 0?
Yes
Acetate is selected! NOTE this and see procedure number
No
Bicarbonate is selected. NOTE this and see procedure number
SECTION 1- FLOW ERRORS IN DIALYSIS
Maximum:
Minimum (no lower than):
23 – 25 psi
18 – 20 psi
Figure
55, page 129).
Figure 55 – CONFIRM Concentrate Screen
Fresenius 2008K Troubleshooting Guide
P/N 507298 Rev. B (Draft 1)
Section 5 – Conductivity/ On Line Clearance Problems
CO- 1.0.2
(page 129).
16 psi
11 psi
CO- 1.0.3
PROGRAM.
(page 130).
CO- 1.0.5
(page 130).
Page 129