The dialysate pressure transducer #9 is bad (see
F- 11.0.1 RESET ALARMS
a) Press and release the 'RESET' key then immediately press and hold it for three seconds. If the "Adjusting
TMP" message is present allow it to expire. If a TMP alarm occurs repeat the RESET sequence once
again then continue.
b) Is ACFS 3.5 or greater?
Yes
See procedure number
No
See procedure number
F- 11.0.3 CHECK CFS ELECTRONICS (ACFS)
a) Referring to the Figure (right), unplug the CFS transducer's
connector (labeled #10) from distribution board position
labeled x10, "CFS".
IMPORTANT!
b)
To avoid time consuming error, using the
screen's clock (upper right), allow up to five minutes (ACFS
response is NOT instantaneous).
c) Does ACFS increase to 9 or greater?
Yes
a) CAREULLY return the CFS transducer #10 to distribution board position "CFS".
b) See procedure number
No
Leaving the CFS transducer unplugged, four possibilities: 1) Bad actuator board
b
board
; 3) Bad sensor cable; 4) Bad distribution board.
a
Swap in a known good actuator board and place the machine into Dialysis program. Call
debug screen 10 and allow up to five minutes. If ACFS goes to greater than 9 the previous
actuator board is bad.
b
Swap in a known good sensor board (see
Options
turns blue). Place the machine into Dialysis program. Call debug screen 10 and allow up to
five minutes. If ACFS = goes to greater than 9 the previous sensor board is bad.
F- 11.0.4 CHECK TMP
Call debug screen 0. Referring to the Figure (right), does the PDlal window read 0 +/- 30?
IMPORTANT!
Yes
a)
the door.
IMPORTANT!
b)
c) See procedure number F- 11.0.5 (page 48).
No
a) ACCURATELY calibrate DIALYSATE PRESSURE. Close the shunt door
when instructed per written procedure in the Calibration book!
b) Return the dialyzer connectors to the shunt and close the door.
c) Connect to concentrate and place the machine into Dialysis program (remember to press the
'CONFIRM' key).
d) See procedure number F- 11.0.5 (page 48).
F- 11.0.1
(page 47).
F- 11.0.4
(page 47).
F- 11.0.3
(page 47).
F- 11.0.4
(page 47).
Hardware Options. Set 'T and C' to "Yes". Press the 'CONFIRM' key (the "x"
Return the red dialyzer connector to the shunt and close
Turn Dialysate Flow ON (Dialysate Flow on/off LED off).
Fresenius 2008K Troubleshooting Guide
P/N 507298 Rev. B (Draft 1)
Section 1 – Flow Errors In Dialysis Program
Figure
70, page 185). Replace it.
Figure
4A, page 8). Enter Service Mode
a
: 2) Bad sensor
Page 47