d) Call debug screen 6 and CAREFULLY watch BC Switch (Balancing Chamber Switching Time) for one
minute. Is it remaining between 324 and 396 (3.24 and 3.96 seconds)?
Yes
The 'Out of Bypass' circuit is okay. See procedure number
No
a) Call the Home screen and check the selected flow rate. It may have defaulted to a
different rate other than 500 ml/min (idle flow option). If (and ONLY if) flow rate has
changed repeat procedure number
b) Proceed to page 14,
TMP- 1.0.7 CHECK 'IN BYPASS' FLOW STABILITY (500 ml/min)
a) Open the shunt door, to cause bypass. Allow 15 seconds.
b) CAREFULLY watch BC Switch for 30 seconds. Is it remaining between 324 and 396?
Yes
The 'In Bypass' circuit is okay. See procedure number
No
a) Call the Home screen and check the selected flow rate. It may have defaulted to a
different rate other than 500 ml/min (idle flow option). If (and ONLY if) flow rate has
changed repeat procedure number
b) Proceed to page 14,
TMP- 1.0.8 CHECK LOADING PRESSURE
a) Close the shunt door.
b) Assuring the loading pressure gauge (with yellow
connector) reads 0 psi before inserting, SLAM it into
the red acetate/Acid rinse port (see Figure right).
c) Per
Table D
(page 179), loading pressure MUST be
fluctuating to its maximum AND
below the minimum. CAREFULLY watch BOTH levels
for two minutes. Is loading pressure okay?
Yes
See procedure number
No
Proceed to page 18, procedure number
Table D – Loading Pressure
®
DiaSafe
Filter?
Yes
No
TMP- 1.0.9 CHECK FILLING PROGRAM FUNCTION (FILACT)
a) Referring to the Figure (right), UNPLUG the air sensor's connector from
distribution board position labeled "AIR-SEN" (4
the left).
b) With "AIR-SEN" empty
continuing! FILACT response in not instantaneous!
c) From debug screen 1, does FILACT = 1?
SECTION 1 - FLOW ERRORS IN DIALYSIS
SECTION 1 - FLOW ERRORS IN DIALYSIS
NEVER
dropping
TMP- 1.0.9
Maximum
No lower than:
23 – 27 psi
15 psi
17 – 23 psi
11 psi
, IMPORTANT!
ALLOW 30 seconds BEFORE
Fresenius 2008K Troubleshooting Guide
P/N 507298 Rev. B (Draft 1)
TMP- 1.0.6
(page 178).
TMP- 1.0.8
TMP- 1.0.6
(page 178).
(page 179).
F-
2.0.4.
th
connector cap from
Section 9 –TMP Problems
TMP- 1.0.7
(page 179).
PROGRAM.
(page 179).
PROGRAM.
Page 179