Rinse or Heat Disinfect: See procedure number
Dialysis: See procedure number
VE- 2.0.2 CHECK VALVE ERROR IN RINSE
a) Place the machine into Rinse Program and allow four minutes.
b) Call debug screen 1. Is VERR greater than 5?
Yes
Assuming that all above procedures were performed correctly, three possibilities: 1) Bad
actuator board; 2) Bad actuator cable; 3) Bad distribution board. To isolate between the
three see procedure number
No
Reseating the connectors may have remedied the Valve Error.
VE- 2.0.3 CHECK VALVE ERROR IN RINSE
a) Place the machine into Dialysis Program (remember to press the 'CONFIRM' key).
b) IMPORTANT! Allow 30 seconds then call debug screen 1. Is VERR greater than 5?
Yes
Assuming all above procedures were performed correctly, three possibilities: 1) Bad
actuator board; 2) Bad actuator cable; 3) Bad distribution board. To isolate between the
three see procedure number
No
Reseating valve connectors may have remedied the Valve Error.
VE- 3.0.0 TROUBLESHOOTING VALVE ERRORS
These procedures check solenoid resistance ( ) from the actuator board, through the actuator cable,
distribution board, and the valve's blue wiring harness:
Check ALL NOTED valves
CAREFULLY read the meter's numeric AND units display (M
ohms; K
= Thousand ohms; "OL" = resistance too big to read)
IMPORTANT!
a)
Turn the machine off!
b) Open the card cage and place a meter lead at chassis ground (see
c) Referring to
Figure 41
(rear) side is arranged in two rows. Odd numbered pins (1, 3, etc) in the bottom row; even numbered
(2, 4, etc) in the top. As instructed
There is NO NEED to check all pins! ONLY the NOTED valves per
VE- 2.0.3
VE- 3.0.0
VE- 3.0.0
(page 86), at the top of the actuator board is a large connector (P2). The solder
later
solenoid resistance(s)
Figure 41 – Electronic Card Cage
Fresenius 2008K Troubleshooting Guide
P/N 507298 Rev. B (Draft 1)
VE- 2.0.2
(page 86).
(page 86).
(page 86).
(page 86).
= Million
Figure
1, page 3).
will be
measured from these pins.
Table 5
Troubleshooting Valve Errors
(page 87).
Page 86