F- 18.0.0 CHECK TMP STABILITY
a) Call the Home screen. Referring to the Figure (right),
the TMP bar graph is populated with three horizontal
bars; yellow upper and lower alarm Limits (positioned
at 520 and below 0) and an 'actual' TMP bar which is
either red (with a TMP alarm) or green (no TMP alarm).
b) Press and release the 'RESET'' key and then
immediately press and hold it for three seconds. If the
"Adjusting TMP" message is present allow it to expire.
If a TMP alarm occurs attempt the RESET sequence
up to twice more then continue.
c) Watch the (green or red) 'actual' TMP bar. TMP is STABLE if (and ONLY if) it REMAINS between the 0
and 520 limits AND varies less than +/- 40 mmHg in two minutes. Proceed accordingly (three scenarios):
1) TMP is UNSTABLE: The 'actual' TMP bar rises ( ) towards the 520 Limit AND/OR falls ( )
towards the 0 Limit, greater than +/- 40 mmHg: See procedure number
2) TMP exceeds the 520 Limit (at some point) causing a "TMP Is High" alarm: See procedure
number
F- 18.1.0
3) TMP is STABLE: See procedure number
F- 18.1.0 TROUBLESHOOT UNSTABLE TMP
NOTE this procedure and page number in case you need to return here:
NOTE
a)
five leak tests will be performed: 1) IOS; 2) 'Spent Side' Balancing Chamber Valves #32,
#34, #36 and #38. See
b) If a leak is not located, see procedure number
F- 19.0.0 CHECK FOR INTERMITTENT "NO WATER"
a) Place the machine into Rinse program for five minutes. Watch CAREFULLY. If (and ONLY if) an
intermittent "No Water" message appears proceed to page 64,
b) Return to Dialysis program (remember to press the 'CONFIRM' key).
c) From the Home screen, set Dialysate flow to 800 ml/min (NOT "a800") and press the 'CONFIRM' key.
d) Watch CAREFULLY watch the screen for five minutes. Does an intermittent "No Water" message appear?
Yes
Proceed to page 64,
No
See procedure number
F- 19.0.1 RINSE AND VERIFY LOADING PRESSURE
Watch loading pressure for three minutes. Does it consistently fluctuate between its maximum and to
LOWER
than its minimum per
Yes
Acid Clean the machine.
No
See procedure number
(page 58).
SECTION 18A – AUTOMATED VALVE LEAK TESTS
SECTION 2 - NO
F- 19.0.1
(page 58).
Table 1
(page 18)?
F- 20.0.0
(page 59).
Fresenius 2008K Troubleshooting Guide
P/N 507298 Rev. B (Draft 1)
Section 1 – Flow Errors In Dialysis Program
F- 19.0.0
(page 58).
F- 19.0.0
(page 58).
SECTION 2 - NO WATER
WATER.
F- 18.1.0
(page 58).
(page 230).
NO
Page 58