Fresenius Medical Care 2008K Troubleshooting Manual page 173

Hemodialysis system
Hide thumbs Also See for 2008K:
Table of Contents

Advertisement

FIL- 3.0.1 CHECK OLC SENSORS
a) Call debug screen 0. Referring the Figure (right), if (and ONLY if)
the Post Dialyzer Cond and Temp sensors do NOT appear skip to
FIL- 3.0.2
(page 171).
b) Are 'pre and post-dialyzer' Temp values within 1.2 of each other?
Yes
See procedure number
No
See procedure number
FIL- 3.0.2 CHECK OLC SENSOR CONNECTORS/ CABLES
Referring to
Figure 65
(page 170), reversed connectors x13 (post dialyzer cond cell) and x44 (post dialyzer
temp sensor) will cause Filling Programs. CAREFULLY trace each cable from the distribution board to the
sensors inside the hydraulics. Check also for insulation damage! Are they reversed or damaged?
Yes
Plug them in correctly (if reversed) or replace them (if the insulation is damaged) then see
procedure number
No
See procedure number
FIL- 3.0.3 CHECK AIR SENSOR #6 ELECTRICAL CONNECTIONS
a) Referring to the Figure (right), assure the air sensor connector is
plugged into distribution board position labeled "AIR-SEN".
b) Referring to
Figure 66
"AIR-SEN" terminates at the male probes on top of chamber #69.
The brown wire should be at the top!
c) From debug screen 1, continue ONLY if FILACT = 1. Reseating
connectors may have remedied the problem!
d) Referring to the Figure (right) and without pulling on the wires
remove both female connectors from the male probes and connect
them snug together. This simulates 'no air in chamber #69'. Leave
them connected together until instructed otherwise!
IMPORTANT!
e)
Allow 30 seconds. FILACT response is NOT
instantaneous.
FIL- 3.0.3
(page 171).
FIL- 3.0.2
(page 171).
FIL- 3.0.8
(page 173).
FIL- 3.0.3
(page 171).
(page 171), assure the wire harness from
Figure 66 – Hydraulics (Front View)
Fresenius 2008K Troubleshooting Guide
P/N 507298 Rev. B (Draft 1)
Section 8 – Filling Program Problems
Page 171