b) Referring to
output nozzle towards the top and the ribbon cable extending from its side. If okay,
replace the bicarbonate pump.
EOS- 4.0.0 CHECK CONCENTRATE PUMP CIRCUIT
a) IMPORTANT! Turn the machine off to prevent damage.
b) Referring to
Figure 63
(i.e. Acid pump to P16, "CONC- P"; Bicarbonate pump to P17, "BIC-P").
c) Five possibilities: 1) Bad actuator board; 2) Bad sensor cable; 3) Bad actuator cable; 4) Bad distribution
board; 5) Bad motherboard.
1. Actuator Board (see
a) IMPORTANT! Turn the machine off and swap in a known good actuator board.
b) Place the machine into Dialysis program (remember to press the 'CONFIRM' key).
c) Allow one minute then call debug screen 0. Is the pump's symbol remaining white?
Yes
Problem solved! The previous actuator board is bad.
No
Continue with EOS- 4.0.0, step #2 (the previous actuator board is okay).
2. Sensor Cable (see
a) IMPORTANT! Turn the machine off and swap in a known good sensor cable (or check
sensor cable).
1
To check the sensor cable. Which pump is issuing the EOS error?
ACID: NOTE that one (1) ACID EOS connection will be checked and proceed to page 228,
SECTION 17- CHECKING THE SENSOR BOARD
BICARB: NOTE that one (1) BICARB EOS connection will be checked and proceed to page
b) Place the machine into Dialysis program (remember to press the 'CONFIRM' key).
c) Allow one minute then call debug screen 0. Is the pump's symbol remaining white?
Yes
Problem solved. The previous sensor cable was bad
No
Continue with EOS- 4.0.0, step #3 (the previous sensor cable is good).
3. Actuator Cable (see
IMPORTANT!
a)
actuator cable).
2
To check the actuator cable. Which pump issuing the EOS error?
ACID: NOTE that four (4) ACID PUMP connections will be checked and proceed to page
225,
cable checks good continue with EOS- 4.0.0, step #4.
BICARB: NOTE that four (4) BICARB PUMP connections will be checked and proceed to
b) Place the machine into Dialysis program (remember to press the 'CONFIRM' key).
c) Allow one minute then call debug screen 0. Is the pump's symbol remaining white?
Figure 63
(page 162), assure the bicarbonate pump is mounted with its
(page 162), CAREFULLY return both pumps to their distribution board positions
Figure
4A, page 8):
Figure
4A, page 8):
good continue with EOS- 4.0.0, step #3.
228,
SECTION 17 - CHECKING THE SENSOR BOARD
cable checks good continue with EOS- 4.0.0, step #3.
Figure
4A, page 8):
Turn the machine off and swap in a known good actuator cable (or check
SECTION 16 - CHECKING THE ACTUATOR BOARD
page 225,
SECTION 16 - CHECKING THE ACTUATOR BOARD
actuator cable checks good continue with EOS- 4.0.0, step #4.
Fresenius 2008K Troubleshooting Guide
P/N 507298 Rev. B (Draft 1)
Section 6 –Concentrate Pump Errors
CABLE. If the sensor cable checks
CABLE. If the sensor
CABLE. If the actuator
CABLE. If the
1
the
2
the
Page 163