AUTOTEST- 2.0.1 REPEAT TEST ON VALVE #31 AND #33 (CHECK UF PUMP)
Leak tests on Valve #43 require the UF pump to achieve pressure during the "Prep" phase. If the UF pump is
malfunctioning other valves that require it will also fail (see
valves #31 and #33 and proceed accordingly:
Both Fail: See procedure number
One or both passes: See procedure number
AUTOTEST- 2.0.2 CHECK UF PUMP
a) Select the screen's 'Calibrate Hydraulics' tab then select the 'UF Pump Volume' tab.
b) Press the 'CONFIRM' key. The screen changes.
c) Place the dialyzer quick connectors into a bucket of water on
the floor and close the shunt door. It's important the
connectors remain submersed!
d) Referring to the Figure (right), remove the Fluid Sample
Connector from its port. No need to attach a burette! The
UF pump is not being calibrated!
e) Press the blinking 'Prime' key. Ignore 5 strokes then watch 5.
Does the UF pump deliver strong pulses during the last 5?
Yes
See procedure number
No
Remove the clamps from valve #43 and proceed to page 215,
AUTOTEST- 2.0.3 CHECK DIALYSATE PRESSURE SYSTEMS
With valve #43 remaining clamped, calibrate DIALYSATE PRESSURE per procedure. Remember to close
the shunt door! Did an "Operator Error" message occur?
Yes
Remove valve #43's clamps then see procedure number
No
Remove valve #43's clamps then repeat
230).
AUTOTEST- 2.0.4 CHECK DIALYSATE PRESSURE SYSTEMS
Remove the clamps from valve #43 and repeat the Leak Test on Valve #43:
Valve #43 fails:
(page 234)
Valve #43 passes: Return to
AUTOTEST- 3.0.0 ISOLATE 'DRAIN CIRCUIT' COMPONENTS
Because valve #30, #29 and the heat exchanger #77 share the 'drain circuit' a leak at any one will cause
valve #29 to fail. The following procedure isolates these components. Four clamps are required:
a) If (and ONLY if) the "IOS" Leak Test was not performed previously do so now. If (and ONLY if) IOS fails
two of three times see procedure number
passed) continue with part b.
b) Referring to
Figure 96
1) Valve #29's clear input tubing.
2) Both sides of the heat exchanger's 'to drain' (CLEAR) tubing.
c) Repeat the Leak Test twice on Valve #29. Does it pass both times?
AUTOTEST- 2.0.2
AUTOTEST- 2.0.3
NOTE
that valve #43 is leaking then see procedure number
B. PERFORM LEAK TESTS
AUTOTEST- 1.0.0
(page 234), isolate valve #30, by tightly double clamp at two locations:
Fresenius 2008K Troubleshooting Guide
P/N 507298 Rev. B (Draft 1)
Section 18A – Automated Valve Leak Tests
Table
29, page 235). Perform Leak Tests on
(page 233).
AUTOTEST- 2.0.4
(page 233).
AUTOTEST- 6.0.0
A. INITIALIZING AUTOMATED LEAK TESTS
(page 230) to repeat the Tests.
(page 231). If IOS already performed (and
(page 233).
UF PUMP
PROBLEMS.
(page 235).
(page
AUTOTEST- 5.0.0
Page 233