OVER- 3.0.0 CHECK INLET WATER PRESSURE (Slow Overflow)
This procedure uses a pressure (psi) gauge to check Inlet Water Pressure Regulator #61. Assure the
gauge reads 0 psi before installing it!
a) Assure incoming water (RO source) pressure is no greater than 105 psi.
b) IMPORTANT! Turn the water off. Referring to
Regulator #61.
Figure 100 – Hydraulics TOP View / Inlet Pressure Regulator #61
IMPORTANT!
c)
Turn the water on!
d) Place the machine into Dialysis program (remember to press the 'CONFIRM' key).
e) Normally pressure fluctuates between a maximum of no greater than 20 psi and to a minimum of no less
maximum
than 8 psi. Is the
Yes
a) Loosen Regulator's #61's lock nut and turn the adjustment screw counterclockwise
until a maximum of between 18 and 20 psi is achieved. If pressure will not adjust
incoming water pressure is too high (greater than 105 psi) OR regulator #61 is bad.
b) Leaving the gauge installed, see procedure number
No
a) Leaving the gauge installed, if pressure remains less than 10 psi ASSURE the water is
on!
b) See procedure number
Figure 99 – Heat Exchanger Check
Figure 100
greater than 20 psi?
OVER- 3.0.1
Fresenius 2008K Troubleshooting Guide
P/N 507298 Rev. B (Draft 1)
Section 20 - Troubleshooting Overflow From The Vent Tube
(page 243), install the gauge at Inlet Pressure
OVER- 3.0.1
(page 244).
(page 244).
Page 243