Fresenius Medical Care 2008K Troubleshooting Manual page 172

Hemodialysis system
Hide thumbs Also See for 2008K:
Table of Contents

Advertisement

IMPORTANT!
e)
With the Bypass LED (front panel) OFF (constant) is the 'bob' inside
the external flow indicator (#75) rising (at least ¼ up) and falling?
Yes
See procedure number
No
a) ASSURE that neither the Temperature nor Conductivity window is RED!
b) If (and ONLY if) Dialysate flow is NOT set to 500 ml/min (NOT "a500") repeat
FIL- 1.0.5
c) Proceed to page 14,
FIL- 2.0.0 CHECK FILACT (Filing Program Active?)
Call debug screen 1 and watch FILACT for one minute.
= 1 (constant or intermittent): See procedure number
= 0 (constant): Filling Program is not occurring. See procedure number
FIL- 3.0.0 CHECK ON-LINE CLEARANCE COMPONENTS
Referring to
Figure 65
(page 170), IF the machine Online Line Clearance (OLC) equipped, both the 'pre-
dialyzer' and 'post-dialyzer' conductivity and temperature sensors are installed. Is the machine OLC
equipped?
Yes
See procedure number
No
See procedure number
FIL- 2.0.0
(page 170).
(page 169) from part e.
SECTION 1- FLOW ERRORS IN DIALYSIS
FIL- 3.0.1
(page 171).
FIL- 3.0.3
(page 171).
Figure 65 – OLC Components
Fresenius 2008K Troubleshooting Guide
P/N 507298 Rev. B (Draft 1)
Section 8 – Filling Program Problems
FIL- 3.0.0
(page 170).
FIL- 5.0.0
PROGRAM.
(page 174).
Page 170