h)
Touch the ohm meter leads together and subtract this resistance ( ) reading from subsequent
resistance readings.
i)
Using a regular meter lead (do NOT use Fluke TP 80), place it on the matching numbered pin (per
Table 26
or
Table
27) on the solder (back) side of the P2 connector. IMPORTANT! Some components
may have more than one pin to check.
j)
Is the meter's numeric AND units display 2
Yes
The cable is good! Return it the distribution board. If referred to
CHECKING THE ACTUATOR BOARD CABLE, return to the procedure
that brought you here.
No
Unplug the cable from the actuator board's P2 connector. Check inside the connector for
bent, missing or damaged male pins. If the pins are okay replace the cable.
Table 26 – Actuator Board, P2 Cable. Pin Identification (for Pumps)
PUMP
DEAERATION PUMP (1 of 2)
DEAERATION PUMP (2 of 2)
FLOW PUMP (1 of 2)
FLOW PUMP (2 of 2)
ACID PUMP (1 of 4)
ACID PUMP (2 of 4)
ACID PUMP (3 of 4)
ACID PUMP (4 of 4)
BICARB PUMP (1 of 4)
BICARB PUMP (2 of 4)
BICARB PUMP (3 of 4)
BICARB PUMP (4 of 4)
UF PUMP (1 of 4)
UF PUMP (2 of 4)
UF PUMP (3 of 4)
UF PUMP (4 of 4)
or less?
PLUG/PIN #
P2 / FEMALE CONNECTOR LOCATION
15
Bottom row, 8 pins from the rear of machine
16
Top row, 8 pins from the rear of machine
19
Bottom row, 10 pins from the rear of machine
20
Top row, 10 pins from the rear of machine
1
Bottom row, first pin from the rear of machine
2
Top row, first pin from the rear of machine
3
Bottom row, 2 pins from the rear of machine
4
Top row, 2 pins from the rear of machine
5
Bottom row, 3 pins from the rear of machine
6
Top row, 3 pins from the rear of machine
7
Bottom row, 4 pins from the rear of machine
8
Top row, 4 pins from the rear of machine
23
Bottom row, 12 pins from the rear of machine
24
Top row, 12 pins from the rear of machine
25
Bottom row, 13 pins from the rear of machine
26
Top row, 13 pins from the rear of machine
Fresenius 2008K Troubleshooting Guide
P/N 507298 Rev. B (Draft 1)
Section 16 – Checking The Actuator Board Cable