NOTE! Valves #24 and #26 locations depend on the On Line Clearance (OLC) option. Valve
#30's location depends on the Aliquot option (see
Figure 35 A – Valve Locations (Hydraulics, Top View)
Figure 35 B – Valve Locations (Hydraulics Rear and Front View)
TSV- 2.0.0 CHECK SOLENOID
a) Referring to
Figure 35 A
locate the NOTED valve(s).
b) Referring to the Figure (right), measure resistance ( ) between the
metal solenoid terminals. Between 40 and 100
Yes
The two-wire (blue) valve wiring harness is bad.
No
Replace the valve.