The Data Number
When the "Data number output (d-no) " is set to "1 " and the data that is stored in the balance memory
is to be output, the "Data number" can be appended just before each data. This format consists of six
characters (excluding the terminator).
Deleting All Data at One Time
11.5.
Step 1
Press and hold the PRINT key until displaying reCall and release the key.
Step 2
Press the RANGE key several times to display Clear .
Step 3
Press the PRINT key to enter this mode.
Step 4
Press the RE-ZERO key. Then the balance displays Clr go .
Step 5
Press the PRINT key to delete all data.
Step 6
The balance displays reCall after the finish.
Step 7
Press the CAL key to return to weighing mode.
12. Underhook
The underhook can be used for weighing large items, magnetic material or density measurement.
The built-in underhook is behind the plastic cap on the under-side of the balance.
Caution
When not using the underhook, attach the plastic cap to prevent dust from getting into the balance.
The underhook can only be used to support items within the weight range of the balance. Do not
overload it.
Operate the underhook gently.
Underhook
Cap
Screw
41