Manual Response Adjustment

6.2.
This way updates the response adjustment manually.
Operation
Step 1
Press and hold the MODE key until RESPONSE is displayed.
Press the MODE key immediately.
Step 2
Select a stage of the response adjustment using the
MODE key. Either FAST or MID. or SLOW can be
selected.
Step 3
The balance automatically returns to the weighing mode after
a few seconds of inactivity.

7. Calibration

The GR series has the following modes concerning calibration and calibration test.
Automatic Self Calibration
Calibration using the internal weight
Calibration using an external weight
Calibration test using the internal weight
Calibration test using an external weight
Correction of the internal weight value
Note
Calibration is controlled by the parameters of "Permission or prohibition". Refer to section "8.
Function Switch and Initialization".
The weight which can be used for calibration is called "the calibration weight". The weight which
can be used for calibration test is called "the target weight". The weight which you have is called
"the external weight".
Caution
This calibration achieves the adjustment for accurate weighing. It is necessary to perform calibration
in the following case.
• When the balance is installed for the first time.
• When the balance has been moved.
• When the ambient environment has changed.
• For periodical calibration.
Prevent vibration, drafts, and ambient temperature changes from the influence for the balance
during calibration.
This indicator means "the balance is measuring calibration data". Do not allow
vibration or drafts to affect the balance while this indicator is displayed.