10.3. Data Output Mode (
)
(Refer to
"12. Function
Settings")
Stream mode (
)
Data is output continuously. Data output is abput 10 times per second.
Command mode (
~ )
The scale is controlled by a command sent from a personal computer etc, connected
externally. For details, refer to "10-4. Command mode".
In
, data is output only by command.
Output by PRINT switch (
)
When the weighing value is stable (stable mark is lit), pressing the PRINT switch will
output the data. The display will disappear for a moment to inform you that data has
been output.
Auto print +/- data (
)
Data is output when the weighing value is stabilized (stability mark is lit) and its value is
+5d or greater, or -5d or smaller. The next output will be after the weighing value has
returned to the range of -4d to +4d.
Auto print + data output (
)
When the value is stabilized (stability mark is lit) and its value is + 5d (d = scale) or
greater, data will be output. The next output will be after the weghing value returns to
+4d or less.
Auto print +/- data and comparison result OK (
)
Data is output when the weighing value is stabilized (stability mark is lit), its value is +5d
or greater or -5d or smaller and the comparison result is OK.
The next output will be after the weighing value has returned to the range of -4d to +4d.
Auto print + data and comparison result OK (
)
Data is output when the weighing value is stabilized (stability mark is lit), its value is +5d
(d = weight minimum display) or greater and the comparison result is OK. The next
output will be after the weighing value returns to +4d or smaller.
21