Explanation of Item "Data output mode"

9.4.
The PRINT key can be used at any time for transmitting data.
Key Mode
When you press the PRINT key and the display value is stable, the balance outputs the weighing data
and the display blinks one time.
Required setting dout
Auto-Print Mode A
When the display value is stable and meets the conditions of "Auto-print polarity", "Auto-print band" and
standard value (of zero point), the balance outputs the weighing data. If you press the PRINT key , the
balance outputs the data and the display blinks one time.
Required setting dout
Example
Auto-Print Mode B
When the display value is stable and meets the conditions of "Auto-print polarity", "Auto-print band" and
standard value (of last stable value), the balance outputs the weighing data. If you press the PRINT key,
the balance outputs the data and the display blinks one time.
Required setting dout
Example
Stream Mode
The balance outputs the weighing data continuously.
Required setting dout
Example
Caution
When the baud rate is set to 2400bps or less, the display update rate is faster than the output rate
and the balance may not transmit the data completely (and transmits it intermittently).
Interval Memory Mode
This is the data memory function mode. Weighing data is periodically stored in the balance. The interval
memory mode can not be used, while stream mode is used.
Required setting dout
Example
prt 0
prt 1
dout
ap-p
dout
ap-b
"The ar-d 1 (zero after output) is set. A sample is placed and removed for weighing
each time."
prt 2
dout
ap-p
dout
ap-b
"Transmitting the data of each operation."
prt 3
dout
data 0
ba5fnc
5pd
5if
bp5
"Monitoring data on a computer"
prt 3
dout
data 1
dout
int
"Periodical weighing without computer command and outputting all of the data to a
computer at one time"
Print key mode
Auto-print mode A
Auto-print polarity
Auto-print band
Auto-print mode B
Auto-print polarity
Auto-print band
Stream mode
Data memory function is not used.
Display update rate
Baud rate
Interval memory mode
Data memory function is used.
Interval time
28