10. ID Number and GLP Report
Main Objectives
10.1.
The ID number is used to identify the balance when Good Laboratory Practice (GLP) is used.
The ID number is output on the "Calibration Report", "Calibration Test Report" and "Title block".
The GLP output format is selected at the "GLP output (info)" of the "Function Table".
The balance can output the following reports for GLP.
"Calibration Report" of the calibration using the internal weight.
"Calibration Report" of the calibration using an external weight.
"Calibration Test Report" of the calibration using the internal weight.
"Calibration Test Report" of the calibration using an external weight.
"Title block" and "End block" for weighing data.
Setting the ID Number
10.2.
Step 1
Press and hold the RANGE key to display
Step 2
Press the RANGE key several times to display id .
Step 3
Press the PRINT key. You can set the ID number using the following keys.
RANGE key
RE-ZERO key
CAL key
PRINT key
Display Character Set
Step 4
With
ba5fnc
The key to increment the digit.
The key to select the character of the digit. Refer to the following table for
the "Display Character Set".
The key to cancel the new ID number and display
The key to store the new ID number and display
displayed, press the CAL key to return to the weighing mode.
ba5fnc .
33
ba5fnc .
ba5fnc .