Data Output Mode (Prt) - AND SJ-3000WP/-BT Instruction Manual

Ip67 compact bench scale
Hide thumbs Also See for SJ-3000WP/-BT:
Table of Contents

Advertisement

Data format
S T
Header
There are three kinds of headers for the weighing value as follows.
ST: Weighing data is stable.
US: Weighing data is not satable.
OL: Data is over.(beyond the measurement range)
Data is always digits including sigh, decimal point.
There are three types of units as follows:
_ k g:
_ _ g:
_ P C:
C
L
is always output for the terminator.
R
F
Example of output data
Weighing data"kg"(+) S T , + 0 0 1 2 . 3 4 5 _ k g
Weighing data "g"(-) S T ,
Overweight

10.3. Data output mode (prt)

Stream mode (prt-0)
Data is output continuously. Data output is abput 10 times per second.
Command mode (prt-0~6)
The scale is controlled by a command sent from a personal computer etc,
connected externally. For details, refer to "10-4. Command mode".
In ptr-1, data is output only by command.
Output by PRINT key (prt-2)
When the weighing value is stable (stable mark is lit), pressing the .PRINT key will
output the data. The display will disappear for a moment to inform you that data has
been output.
Auto print + data output (prt-3)
When the value is stabilized (stability mark is lit) and its value is + 5d (d=scale) or
greater, data will be output. The next output will be after the weghing value returns
to +4d or less.
Auto print +/- data (prt-4)
,
+ 0 0 0 0 0
Data
Unit amount of weighing data"kg"
Unit amount of weighing data "g"
Unit amount of number
"kg"(+) O L , + 9 9 9 9 . 9 9 9 _ k g
.
0 0 _ k g
Unit
"PCS
- 0 0 0 0 1 2 3 4 _ _ g
19
C
L
R
F
Terminator
"
C
L
R
F
C
L
R
F
C
L
R
F

Advertisement

Table of Contents
loading

Table of Contents