Waking Up the Battery With a Monitor

How to Recycle the Battery

Revision A
Maintenance: Lithium-Ion Battery
1. Connect the monitor to an AC power source.
2. Activate the Boot Loader program.
a. Hold down NBP GO/STOP and ZERO ALL.
b. Press and release the Trim Knob control.
c.
Keep holding NBP GO/STOP and ZERO ALL until the Boot
Code information appears on the display.
3. Select Wake Up Battery from the Service Menu.
4. Insert the "sleeping" battery into Slot A of the monitor.
5. Select Wake Up Lithium Ion Battery in SLOT A. Then, the monitor
applies a "wake-up" shot of current for up to 210 seconds.
6. Watch for the message, Battery awake, run main code to charge. The
battery should now be "awake."
If the "Battery awake, run main code to charge" message appears
but the battery is NOT awake, the battery may be in a deep
discharge state. You should repeat step 5 at least four more times
to give the battery additional "wake-up" shots of current.
7. Quickly reboot the monitor to begin charging the batteries. The
monitor's illuminated yellow-colored CHARGING STATUS LED
indicates that the battery is being charged.
127(
The monitor will not charge the battery while it is running the
Boot Loader program.
When the battery no longer holds a charge, it should be replaced. The
battery is recycleable. Remove the old battery from the monitor and
follow your local recycling guidelines.
:$51,1*
EXPLOSION HAZARD — DO NOT incinerate the
battery or store at high temperatures. Serious injury or
death could result.