Camera, Coarse Focus, Scan Width & Tilt Adjustment
For all cameras:
1. Open the Original Guide and place OAP in the scan area. Make sure to position the pattern by
pushing it fully forwards and to the right.
2. Deselect RGB.
3. Select Normal.
4. Loosen the Focus Lock Screws (3).
5. Adjust the focus by turning the Focus Adjustment Screw (4) until focus is as good as possible.
6. Select Camera Adj.
7. Lightly loosen the Camera Base Fixing Screws (1).
8. Select camera A (default).
9. Adjust the Scan Width Adjustment Screw (2) until the arrow in the Scan Width box (circled
with red in image below) is within the black box (arrow turned green).
10. Repeat for other cameras (select B, C, D, E in turn). Tighten the Camera Base Plate Fixing
Screws (1).