55-509
IT8 file is not accessible.
If you have received a new Calibration Sheet, please allow application to
find reference file over the internet.
If problem persists reinstall the software to correct the issue.
55-513
The IT8 reference file could not be found on the internet
Browse to it manually www.contex.com and retrieve the correct IT8 file or
use standard
55-523
Sheet not recognized.
Please clean scanner.
Please run application again.
Inspect the sheet for scratches or wear, and replace it if the problem
continues.
55-530
When checking new calibration the result was not within the limits.
Inspect the sheet for scratches or wear, and replace it if the problem
continues.
55-611
Unknown Scanner Status Error.
Reboot system.
55-613
No paper was detected in the scanner.
Place the correct sheet in the scanner.