Error code:
Keypad error code:
Level:
Description:
Potential fail comp:
Corrective Action:
Error code:
Keypad error code:
Level:
Description:
Potential fail comp:
Corrective Action:
Error code:
Keypad error code:
Level:
Description:
Potential fail comp:
Corrective Action:
Error code:
Keypad error code:
Level:
Description:
Potential fail comp:
Corrective Action:
Error code:
Keypad error code:
Level:
Description:
Potential fail comp:
Corrective Action:
100-50231
BOOT
RD Correctable
Firmware download is in progress
Firmware
Firmware.
100-50234
NONE
User Correctable
Scanner is in safe-mode.
User; SUx
User interaction.
Please reboot scanner or download new firmware.
Main controller board (SUx).
100-50237
ERR_GENERAL
User Correctable
Host PC does not support SSE2 calculations.
User
User interaction.
100-50262
ERR_GENERAL
Support Correctable
LMx, Unable to communicate with LMx board.
LMx; SUx; Cable
Lamp and Stepper Motor driver board (LMx)
Main controller board (SUx).
Related cabling.
100-20088
ERR_GENERAL
Support Correctable
Stepper Motor system error; motor movement not detected.
Mech, SUx, Cable, Stepper Motor
Belt & Stepper Motor
Main controller board (SUx)
HD Ultra X Service Guide